Description and Specification for Lab Tenders
Horizontal High Speed Cylindrical Steam Sterilizer.
Introduction
The sterilizer is based on the principle that microbiological action
of saturated steam at elevated temperature is rapid and through. All
models work on the principle of downward displacement of air which is
the most economical method of obtaining sterilization.
Robust and rigid construction, designed for all types of bulk
sterilization which are commonly needed in Medical, Agricultural and
Pharmaceutical Institutions.
The high speed sterilizer is fitted with high wattage heaters. Primarily it is intended for emergency sterilization of unwrapped surgical instruments & other components.
Construction
The sterilizer is triple walled with steam jacket and separate boiler.
The inner chamber and steam jacket are made up of Heavy gauge
Stainless Steel sheet with leak proof argon welding. The sterilizer
has single piece door made of stainless steel plate. Back plate and
ring is also made of thick stainless steel. All the sterilizers are
hydraulically tested to withstand 2.5 times the working pressure.
The sterilizer is mounted on tubular steel frame with ground levelling
screwed flanges. The Outer jacket is wrapped with glass wool to
minimise the heat losses due to radiation and is covered by polished
stainless steel sheet for that elegant appearance.
Steam Generator
The steam generator is made of heavy stainless steel sheet. Heavy ring
mounted in front of the boiler with removable thick stainless steel
plate is fitted with heating elements and low water level cut-off
device to protect these elements from burning out dry.
Front removable plate system provides easy cleaning of the deposited
scale on the elements and the walls for long life and efficiency. The
boiler is fitted with water gauge glass for water level indication,
water inlet and outlet valves.
Pressure Control
The “PIEZOSTAT” (Pressure control device) is incorporated in all
sterilizers. It acts as a safety device, economises on power
consumption and reduces the frequent opening of the steam release valve
and thus prevents the release of steam in the room, It cuts off the
power supply to heaters when the set pressure is achieved and
re-energises the heating elements when pressure falls below the set
point.
Heating Elements
Flanged types, immersion heating elements made of high grade material
are high wattage electric heaters to finish the sterilization within 6
minutes of achieving the desired pressure of 2.2Kg/cm2 & corresponding saturated steam temperature of 134°C.
Safety Door
Sterilizer has hinged self-locking single piece thick Stainless Steel
plate door and cannot be opened when under pressure. The self-locking
device automatically disengages the thread mechanism when the
sterilizer is under pressure to preclude any eventuality of accident.
Single Point Control
Complete sterilizing cycle is controlled from one point with the help
of “Multiport Valve” fitted at front top. Two speeds of steam exhaust
are available; fast and slow.
Easily readable jacket and chamber pressure on gauges mounted on the
left and right of the multi port valve. A moisture trap fitted in the
chamber discharge line absorbs the condensate automatically to prevent
moistening of the subject matter to be sterilized.
Control Panel:
The control panel includes a pressure switch, a microprocessor based
digital temperature indicator cum controller, Mains MCB, pilot lamps,
relays etc. and mounted on the stand of the sterilizer for easy
operation.
Dia × Depth |
Load |
400 × 600 mm |
18 kw |
500 × 900 mm |
24 kw |
Categories: Sterilization Equipment
AMBALA, India